7 Tests and measurements

See chapter 6.1 Selection modes for instructions on keys and touch screen functionality.

Voltage, frequency and phase sequence

7.1
Measurement parameters
1)
System
Limit type
Earthing system
2)
Nominal voltage
1)
There are no limits to set if System parameter is set to '–'.
2)
Active only if limit type is set to %
Refer to chapter 4.6.5 Settings for more information.
Measurement limits for TN/TT earthing system:
3)
Low limit Uln
3)
High limit Uln
4)
Low limit Uln
4)
High limit Uln
3,4)
Low limit Ulpe
3,4)
High limit Ulpe
3,4)
Low limit Unpe
3,4)
High limit Unpe
5)
Low limit U12
5)
High limit U12
5)
Low limit U13
5)
High limit U13
5)
Low limit U23
5)
High limit U23
6)
Low limit Ull
6)
High limit Ull
3)
In case of 1-phase voltage system and limit type set to voltage.
4)
In case of 1-phase voltage system and limit type set to %.
5)
In case of 3-phase voltage system and limit type set to voltage.
6)
In case of 3-phase voltage system and limit type set to %.
Figure 7.1: Voltage measurement menu
Voltage system [-, 1-phase,3-phase]
Type of limit [Voltage, %]
Earthing system [TN/TT, IT]
[110V, 115V, 190V, 200V, 220V, 230V, 240V, 380V, 400V, 415V]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [-20% ... 20%]
Max. voltage [-20% ... 20%]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [0 V ... 499 V]
Max. voltage [0 V ... 499 V]
Min. voltage [-20% ... 20%]
Max. voltage [-20% ... 20%]
90
Tests and measurements