MRM Malibu/Sovereign
Trouble shooting
Before starting to trouble shoot check:
Power supply, Water supply, LED:s on Panel and in Control box and emergency stop
Problem
Nothing happens
when pressing any
button (no LED:s)
Temperature display
tub shows room tem-
perature when filling
hot water
Note: No el. scalding
protection
Display fill /shower
shows:
--.-
when NOT
activated (Tempera-
ture display tub work-
ing) Note: No el.
scalding protection
shower.
Both temperature dis-
plays black?
Note: No el. scalding
protection.
-LED 5 in Normal Mode shows if HI/LO actuators are in lowest position
Tub will not rise
Tub is not rising par-
allel
Cause
CONTROL PANEL
Overlay defect, no
connection J16 in
control box
Bad connection
between sensor and
tub nipple
Shortcut, no connec-
tion of fill/shower
cabling/sensor or no
connection from con-
trol box
No connection/ ther-
mometer defect
HI/LO
Lift arm outside tub
Electric fault
Actuator position has
been changed
Action
Check control box
Replace control panel
Apply contact paste
Check sensor/cable to J11
(fill in) or
Check sensor/cable to J12
(shower in)
Check cable to J10
Replace thermometer
Check connection J8 in
Control box/replace ther-
mometer
Swing the lift arm into the
tub
Check hand control and
cables to J15
Check Hi/Lo actuators and
cables to J21,22,23,24
Run actuators to lowest
position to reset to zero
53
Trouble shooting
Service
procedure
See page 48
No fault Refer
to Operating
Instructions
SP 26