Service Procedures
5. Unscrew the PEX pipes to the mixer.
NOTE
Hot water may come out of the mixer
6. Slightly loosen the two screws (a) holding
the mixer to the u-bar and remove the screw
(b) holding the bracket (e) to the chassis
7. Pull out the mixer
8. The mixer shaft (c) will pop out of the mixer
hub (d) and can be left hanging.
Removing the adjustment plate and
cartridges
• Remove the four screws (f) holding the
adjustment plate (g).
NOTE
The cartridges (h) are spring loaded (i) and can
pop up very hard when all screws are removed.
Only cartridges with a white top should be used.
• Use repair kit with three cartridges -
replacing single cartridges or o-rings will
not restore to full performance
• Replace check valves (j) and o-rings (k)
included in kit
-Position of check valve shall be according
to picture on bottom of this page.
-Check that no remains of the old check
valve is left inside.
• Lubricate all o-rings
Assembling
1. Tighten the screws (f) in diagonal order to
make the cartridges fit correctly
2. Hold the mixer in place and position the
mixer shaft (c)
3. Attach an tighten screws holding mixer (a,b)
4. Attach PEX pipes
5. NOTE: Adjust the mechanical scalding
protection (see page 16)
6. Test function
7. Attach inner cover and front panel
c
b
Mixer
Cartridges and adjustment plate
14
MRM Malibu/Sovereign
d
e
a
f
g
h
i
j
k