To test the emergency stop:
1. With mains on - push and release the stop
button (a) a few times to be sure the tub func-
tions are interrupted.
WARNING
When the mains is shut off by the emergency
stop there is still power connected in the con-
trol box. That's why the mains indicator on
the control panel still shows power on with
the green LED.
2. Check that the emergency stop button is
securely fit to the tub. A rotation stop washer
is used here since september 2010 (b).
3. If no function check the emergency stop with
connection (connected to the control box
position J4) Replace if not functioning.