Service Procedures
the technician must state in the service notes
that the castors will remain safe until the
next scheduled service.
NOTE
All castors must be of the same brand. It´s not
permitted to use different versions of castors
on the same product.
Replacement
1. Unscrew the castor (c) including washer (d)
2. Remove old Loctite (Refer to Loctite
instructions)
3. Apply new Loctite (type see page 47)
4. Assemble the new castor with bolt and
washer (torque see page 47).
5. Check the function of the new castor.
32
MRM Malibu/Sovereign
c
d