Check/Replace Flow Restrictors - Arjohuntleigh Malibu Sovereign Maintenance And Repair Manual

For internal use only!
Table of Contents

Advertisement

Service Procedures
SP 7. Check Flow Restrictors
If the water regulation of the MPA is disturbed
check:
1. That the flow and pressure of incoming wa-
ter is sufficient.
2. If the flow and pressure of the incoming
water is OK then measure the flow from the
shower, (default 4,9-5,7 l/min) by filling a
vessel with known volume
3. Calculate the flow by dividing the volume of
the vessel with the time to fill the complete
vessel e.g. 20 l /4 min=5 l/min
4. Perform the test twice
5. If the average flow is 4-5 l/min or more there
is no need to replace the flow restrictors
6. If the average flow is less than 4 l/min both
the fill valve (a) and the pump valve (b)
must be replaced, follow the replacement
procedure below.
Replacement procedure
1. Shut off water and mains
2. The fill valve is reached in the front of the
tub while the pump valve is reached from
the side of the tub
3. Disconnect the hoses from the fill valve and
unscrew it from the holding plate. When
assembling check the cable connection
according to picture top right and fit the
hoses
4. Pump valve, bottom view right:
Disconnect the hoses from the valve and
unscrew it from the hose nipple (c).
Note. With auto flush cabling for valve 1 (d)
is not connected and line (e) is plugged.
5. When assembling the pump valve, note
positioning; back outwards
Red
Green
Orange
Black
c
d
Cable connections for 3-way valves
12
MRM Malibu/Sovereign
a)Fill valve
b)Pump valve
Yellow
Brown
3
2
1
Front view
Brown
Yellow
2
3
Green
1
e
Orange
Black
Back view
Red