Service Procedures
SP 10. Check Shower/
Disinfection
Handle
Shower handle
Check shower handle for bad flow, leaks/
damage and replace if necessary.
1. Turn the adjuster ring (a), fully anti clock-
wise
2. Unclip and remove the adjuster ring (a)
3. Unscrew the spray plate (b) anti clockwise.
The o-rings (c) will provide some resistance
4. Clean parts (a,b,c) with a stiff brush in water
or if necessary delime
5. Assemble the o-rings (c) and make sure they
are located correctly
6. Turn the flow divertor (d) fully clockwise
7. Carefully screw the spray plate (b) onto the
flow divertor (d). There should be a gap
between the edge of the spray plate and the
flow divertor
8. Make sure the green dot (e) is according to
the picture and align inner teeth (f) with
outer teeth (g)
9. Clip the adjuster ring (a) in position.
With auto flush shower holder
• Check handle holder for cracks, leaks and/or
sharp edges
• Check fastening of handle holder
• Repeat insertion and removal of shower
handle into holder for safe fit
Disinfection handle (Yellow)
Note, grey shower handle is illustrated.
Repair kit A:
1. Knock out two taps (a) holding trigger and
remove the trigger, there is a sleeve inside
2. Pull out the plastic stop (b)
b
b
19
MRM Malibu/Sovereign
g
d
c
a
a
e
f