MRM Malibu/Sovereign
SP 9. Check Dosage
Disinfection
Perform the procedure in a room temperature
between 20-25°C (68-77°F) to obtain correct
set values.
1. Open the panel lid and remove the disinfec-
tion bottle
2. Fill a graduated 50 ml cylinder with your
used disinfection fluid.
3. Place the cylinder where the disinfection
bottle was placed
4. Place the bottle connection (a) in the
cylinder
5. Hold the disinfection handle in tub rim level
and pointing downwards
6. Start the disinfection by pressing the
disinfection button (b) for 5 seconds
7. Press the handle trigger to get a disinfection
flow, and keep it activated until the flow is
stable and you no longer have any air
bubbles in the fluid.
8. Check the set value on the flow meter (c)
according to the recommended value (see
table below), release the handle trigger
Disinfection agent
ArjoClean up to
VEPL00005449
ArjoClean from
VEPL00005450
9. Top up the graduated cylinder completely,
without removing the bottle connection
10.Use a graduated 2000 ml container to collect
the disinfection mixture
11.Start the disinfection, remember to hold the
handle in tub rim level
12.When the level in the 50ml cylinder passes
50 ml start to spray the disinfection mixture
into the container until the level reaches 20
ml
13.Shut off the disinfection by pressing the
disinfection button (b)
b
Value
ml/l
70
16
50
16
17
Service Procedures