Troubleshooting 2200 TapeStation ScreenPlex Images

Table 8
Overview troubleshooting 2200 TapeStation images
Image
Issue
Only the top and
bottom markers are
present in the
sample.
The bands appear
diagonally across the
lane.
2200 TapeStation

Troubleshooting 2200 TapeStation ScreenPlex Images

Probable cause
Solution
Either the sample is
Always vortex samples for 5 seconds before
not present or is too
placing in the 2200 TapeStation. ScreenTape can
faint to visualise.
detect <1 ng/µL
A bubble was located
Always flick the ScreenTape before placing in the
on top of the gel at
2200 TapeStation.
the gel/buffer
interface.
6
Troubleshooting
57