Maintenance

General Information

General Information
Annual Preventative Maintenance (PM) is essential for the TapeStation as it
has many moving parts.
This PM should be arranged with your local Agilent representative and
consists of
• Fan Filter replacement
• Needle replacement
• Electrophoresis probe replacement
• Internal instrument inspection for wear, foreign objects and general clean
In addition to the above, the engineer will check that the instrument is
functional by running a full ScreenTape.
For customers with exceptionally high usage the needle replacement
procedure as detailed in the section below can also be performed between
annual PM services.