Agilent Technologies 2200 TapeStation User Manual page 63

Hide thumbs Also See for 2200 TapeStation:
Table of Contents

Advertisement

Table 9
Overview troubleshooting test kit manufacturers TestKitSuite.xml file
Image
Issue
A warning sign has
appeared in the
result table next to
an expected
fragment and a
corresponding
warning sign has
appeared next to a
fragment in the
image.
An unidentified band
has also been
reported in the table,
but not in the gel.
A warning sign has
appeared in the
result table next to
an expected
fragment and a
corresponding
warning sign has
appeared next to a
fragment in the
image.
2200 TapeStation
Troubleshooting the Test Kit Manufacturers TestKitSuite.xml File
Probable cause
The warning signs in the result table indicate that
the internal control (marked as important in the
TestKitSuite file) is missing and that an
unexpected, and therefore unidentified, band has
been detected.
The gel image actually contains 2 warning signs
but they are very close together and therefore
cannot be distinguished.
Therefore, in this example, the internal control
has not been identified within the expected
diagnostic window set in the
file but may have been identified out with any
predetermined window, creating an unidentified
fragment.
In this example, the warning sign in the result
table indicates that the internal control is missing.
In the gel image, warning sign indicates that the
internal control diagnostic window is empty.
However, the internal control has not been
marked as unidentified. Therefore, it may have
been sized within an ignore region marked in the
TestKitSuite and has not been reported
Troubleshooting
Solution
Run the sample
again. If the issue
re-occurs, contact
the Test Kit
Manufacturer.
TestKitSuite.xml
Run the sample
again. If the issue
re-occurs, contact
the Test Kit
Manufacturer.
6
63