Using the 2200 TapeStation System
Operating Procedure
The ladder must be selected first to ensure it's loaded into the first available lane.
N O T E
ScreenTape can be used up to 2 weeks after first use if it has been stored upright between
N O T E
2 – 8 °C.
Simply select the samples in the same manner as whole ScreenTapes. The first sample
selected will automatically appear in the first available lane and will be annotated as the
ladder.
10 Once the samples have been selected, a menu will appear for the test
11 If one test is to be applied to the whole tape, then all the wells should be
12 Selected but unused ScreenTape lanes will be annotated with (Blank Test) in
Pressing Escape on the keyboard will also cancel the current selections.
N O T E
30
selection.
selected at the same time. However, if you wish to run several tests on one
tape, then each test's samples should be selected separately.
the Test Kit column of the sample description table.
The sample selection and corresponding test can be deleted by right
clicking on the sample plate image. A menu will appear with the following
options:
1 Clear All Selections – this will clear ladder well and all sample wells
selected
a Clear Last Selection – this will only clear the samples highlighted by the
user for the last test selected.