Contents Contents 1 Introduction to the 2200 TapeStation System Overview of the System 2 Specifications Technical Specifications 3 Installing the System Unpacking the System Contents of the 2200 ScreenPlex TapeStation Installing the System 4 Using the 2200 TapeStation System Intended Use of the 2200 TapeStation System...
2200 TapeStation Introduction to the 2200 TapeStation System Overview of the System ScreenTape Architecture Agilent 2200 TapeStation Components This chapter gives an introduction to the system Agilent Technologies...
(ScreenTape), an instrument (the 2200 TapeStation) and an analysis software. The system is very straightforward to use, simply place the sample tubes and ScreenTape in to the 2200 TapeStation and let it load, separate, image, analyse and present the results.
Page 7
Introduction to the 2200 TapeStation System Overview of the System Information pertaining to the 2200 TapeStation can be found in: • 2200 TapeStation Technical Specification • 2200 TapeStation Components • Installing the 2200 TapeStation Information pertaining to sample and ScreenTape requirements can be found •...
USB Socket: The USB connector is inserted into the USB socket to link the laptop to the 2200 TapeStation. Power-cable socket: The power cable must be connected to the 2200 TapeStation and the relevant mains electricity outlet. 2200 TapeStation...
Page 9
Introduction to the 2200 TapeStation System Overview of the System ScreenTape Tip holder Sample block Figure 2 2200 TapeStation (front view) On/off switch Power-cable socket USB socket Figure 3 2200 TapeStation (back view) 2200 TapeStation...
Page 10
Introduction to the 2200 TapeStation System Overview of the System 2200 TapeStation...
Page 11
2200 TapeStation Specifications Technical Specifications This chapter provides information on the specifications of the 2200 TapeStation. Agilent Technologies...
Specifications Technical Specifications Technical Specifications 2200 TapeStation Input voltage: 12 V DC Power consumption: 40 W (VA) Current: Interface: USB cable (PC comms.) Instrument Housing: UL94/VO rated flame retardant cast polyurethane Dimensions: 400 x 310 x 310 mm Weight: 12.5 kg...
Page 13
Check that any air conditioning is functioning. • If the quoted current temperature is below the specified range please allow the instrument to equilibrate to the ambient temperature, and avoid using in a cooled area. 2200 TapeStation...
2200 TapeStation Installing the System Unpacking the System Contents of the 2200 ScreenPlex TapeStation The TapeStation ScreenTape Products Installing the System Software Installation Agilent 2200 TapeStation Set Up This chapter gives information about how to install the system. Agilent Technologies...
Unpacking the System Unpacking the System Unpacking the Agilent 2200 TapeStation Do not attempt to unpack the 2200 TapeStation instrument until you have read the accom- panying Site and Safety Manual. Condensation within the instrument C A U T I O N Condensation will damage the system electronics.
Page 17
Ensure all parts and materials have been delivered with your system. The delivery checklist is shown below. Please report any missing or damaged parts to your local Agilent Technologies sales and service office. Please note that ScreenTape and Reagent kits are not provided with instrument delivery N O T E and must be purchased separately.
Installing the System Contents of the 2200 ScreenPlex TapeStation Contents of the 2200 ScreenPlex TapeStation The TapeStation Table 1 Agilent 2200 TapeStation ScreenPlex System (G2966AA) Product Volume Properties Agilent 2200 The instrument for loading, electrophoresing, imaging and analysing double TapeStation...
Installing the System Installing the System Software Installation The software for your Agilent 2200 TapeStation system is preinstalled on the system laptop. For updates, or if you have to change the laptop, you may download the latest version of the N O T E software from the Test Kit Manufacturer's website.
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. For general safety information, please refer to the 2200 TapeStation System - Site and N O T E Safety Manual.
Page 22
As there is more than one driver that can be detected and installed, you may need to follow N O T E these steps more than once. You may need to follow these steps if you change the USB port on the laptop for the TapeStation connector cable. 2200 TapeStation...
Aborting the TapeStation Run Complete TapeStation Run Empty Tip Buckets How to Use the ScreenPlex Analysis Software Shutdown and Restarting Procedure How to prepare your samples DNA Sample Preparation This chapter explains the intended use of the 2200 TapeStation System. Agilent Technologies...
1 – 30 ng per fragment with a maximum concentration for one test sample of 75 ng. The sizing reproducibility for the same double stranded DNA species is within ±10 %. The 2200 TapeStation ScreenPlex System is intended for professional use only. Results obtained using this system must be interpreted in conjunction with other relevant data.
Using the 2200 TapeStation System Performance Limitations of Use Performance Limitations of Use The 2200 TapeStation Screenplex System can analyse a maximum of 15 samples at any one time. The user is responsible for establishing performance characteristics necessary for upstream and downstream applications. Appropriate controls must be...
Available tests • Thresholds required for peak detection • Fragment size ranges required for positive and negative readout) • 2200 TapeStation System (Agilent 2200 ScreenPlex Software) Components • ScreenTape DS12, Ladder and Loading Buffer (stains and separates amplified DNA) •...
Using the 2200 TapeStation System Operating Procedure Operating Procedure 1 Double click the 2200 TapeStation controller icon on the desktop and follow the instructions on the screen. Always ensure you are using the most up-to-date Controller software. Please check for the N O T E latest version.
Page 28
If any used tips are left in the tip-buckets, a pop up window will ask for the discarded tips to N O T E be removed. The 2200 TapeStation will not run until all the tip buckets are empty. A Loading tip transfer tool (G2964-60000) is available.
Page 29
39 or the appropriate Quick Guide. 8 Place samples into the sample block inside the TapeStation. Damage to the 2200 TapeStation and impact on performance C A U T I O N ➔ Ensure the lids have been removed from the sample tubes.
Page 30
Using the 2200 TapeStation System Operating Procedure The ladder must be selected first to ensure it's loaded into the first available lane. N O T E ScreenTape can be used up to 2 weeks after first use if it has been stored upright between N O T E 2 –...
Using the 2200 TapeStation System Operating Procedure Add Experiment Notes 1 If required, notes can be manually entered into the software before the instrument is started. This information will appear in the analysed results. N O T E 2200 TapeStation...
Using the 2200 TapeStation System Operating Procedure Describe Samples 1 Sample descriptions can be manually entered into the software before the instrument is started and whilst the TapeStation is operating, before analysis software is launched. If samples are barcoded, place the cursor in the desired track description and scan the sample barcode with a hand-held barcode scanner.
Using the 2200 TapeStation System Operating Procedure Start the TapeStation Run 1 Click the start button. This will produce a Save As window. As a default the file name starts with the date, in reverse order, and a run counter. When run continuously, the save function auto increments the counter part of the file name.
Using the 2200 TapeStation System Operating Procedure Running System Exposure to potentially dangerous mechanical parts WA R N I N G ➔ Do not open the lid whilst the light is flashing. Aborting the TapeStation Run 1 If, for any reason, you wish to abort an experiment, click the abort button on the pop-up controller.
Using the 2200 TapeStation System Operating Procedure Empty Tip Buckets 1 Empty tip buckets. Used loading tips must be removed from the tip buckets before the next experimental run. N O T E The TapeStation will not start if tips are detected in the buckets.
Using the 2200 TapeStation System Operating Procedure How to Use the ScreenPlex Analysis Software Once the imaging is complete, the analysis software will automatically launch the ScreenPlex Result File (extension .srf). The data provided includes the test types selected (displayed in tabs), the date and time at which the samples were run, the tape serial number and the system information.
Page 37
Using the 2200 TapeStation System Operating Procedure annotated with a warning sign on the image. If the internal control is undetected, a warning sign will illustrate the expected position on the gel image. • The review pane is accessed through Review in the top right hand corner of the ScreenPlex Result Viewer.
Using the 2200 TapeStation System Operating Procedure • Clicking Reset All will remove all manual changes and return the entire tape to the automatically analysed state. • A text file, raw gel image (bitmap), Word and Excel report can all be exported and saved for each run.
Damage to the 2200 TapeStation instrument C A U T I O N ➔ Use only the recommended consumables and reagents with the 2200 TapeStation system. DNA ScreenTape, Ladders and Loading Buffers should be stored vertically between 2 – 8 °C N O T E in their original unopened packaging.
Page 40
For successful loading, the sample solution must be placed at the bottom of the tube or N O T E well without any air-bubbles. The 2200 TapeStation will load a sample from a minimum of 3 µL onto ScreenTape. 2200 TapeStation...
The Ladder must be used with every instrument run and selected to load into the first N O T E available ScreenTape lane. Only Agilent Technologies provided ladder and loading buffers must be used with the N O T E Agilent 2200 TapeStation ScreenPlex System.
Page 42
Using the 2200 TapeStation System DNA Sample Preparation 2200 TapeStation...
2200 TapeStation Maintenance General Information Changing the Needle This chapter describes the maintenance of the TapeStation system. Agilent Technologies...
In addition to the above, the engineer will check that the instrument is functional by running a full ScreenTape. For customers with exceptionally high usage the needle replacement procedure as detailed in the section below can also be performed between annual PM services. 2200 TapeStation...
TapeStation for Nucleic acids Twin G2960-60063 ST010 TapeStation for Protein / Combined TapeStation G2960A 2200 TapeStation System G2961A 2200 TapeStation Nucleic Acid System G2964AA 2200 TapeStation System Twin G2960-60063 G2965AA 2200 TapeStation Nucleic Acid System G2966AA 2200 TapeStation ScreenPlex System...
Page 46
• After 4480 pierces (8960 lanes in a Dual loading system), the needle has completed its lifetime and must be changed before the TapeStation will start. Figure 5 Controller software indicating a Needle change is recommended 2200 TapeStation...
Page 47
Needle cartridge (for use in dual pump systems) For use with product numbers ST017, ST019, ST010, G2960A, G2961A, G2964AA, G2965AA and G2966AA New needles cartridges can be ordered at any time from Agilent Technologies by N O T E contacting your local sales agent.
Page 48
Maintenance Changing the Needle 2200 TapeStation...
A ScreenTape has The barcode has not Power cycle the 2200 TapeStation and restart the been inserted but been scanned by the laptop. If the problem persists, contact the TestKit has not registered.
Page 51
The experimental The controller has Manually type in the desired experiment number – number has started been shut down and the 2200 TapeStation will then restart at 01 again. restarted. automatically incrementing. 2200 TapeStation...
Agilent 2200 ScreenPlex Software Error Messages Message Probable cause Solution Please power cycle An error has Switch the 2200 TapeStation off and then back on the instrument occurred during the again. experimental run. Failed to connect to The drivers have not Power cycle the 2200 TapeStation.
This may occur over Holding the tape with the ScreenTape label facing appeared in the time. you, flick the bottom of the tape. This will move buffer chamber of the any bubbles to the top of the chamber. tape. 2200 TapeStation...
2200 TapeStation before opening the lid. If the LED is constant, the lid can be opened as the 2200 TapeStation is idle. If the LED is flashing, do not open the lid as the 2200 TapeStation is running. Tips are stuck on The wrong tips may...
5 seconds and spun down the 2200 TapeStation optimally. before placing in the 2200 TapeStation. has provided a different result to the previous run. A completely blank There may be a...
Page 56
Ensure you check the top of the gel for bubbles sample fragment a bubble at the top of prior to running in the 2200 TapeStation. Any bands in the gel the gel. bubbles present at the top of the gel can easily be image appear slanted moved out of the way.
Only the top and Either the sample is Always vortex samples for 5 seconds before bottom markers are not present or is too placing in the 2200 TapeStation. ScreenTape can present in the faint to visualise. detect <1 ng/µL sample.
Page 58
The lane has not Always spin down samples before placing in the and/or ladder bands loaded optimally. 2200 TapeStation to remove any bubbles. appear very faint. Very rarely, the 2200 TapeStation does not load effectively - run the sample again. 2200 TapeStation...
Page 59
Troubleshooting Troubleshooting 2200 TapeStation ScreenPlex Images Table 8 Overview troubleshooting 2200 TapeStation images Image Issue Probable cause Solution A mark has appeared Dirt or specks of dust Deselect the annotated ghost band or run the lane in the lane which has can adhere to the again.
Troubleshooting the Test Kit Manufacturers TestKitSuite.xml File Troubleshooting the Test Kit Manufacturers TestKitSuite.xml File Agilent Technologies accepts no responsibility for the contents of the TestKitSuite.xml file. N O T E Please contact the Test Kit Manufacturer for any troubleshooting advice that cannot be located below.
Page 61
The detection of any unexpected fragments that are sized out with these predetermined windows are marked as unidentified and presented with a warning sign in the gel and result table. 2200 TapeStation...
Page 62
Manufacturer. This is usually used for known non-specific have a pathogen amplification and/or primer dimers. marker sign next to them and are not reported in the result table. 2200 TapeStation...
Page 63
However, the internal control has not been Manufacturer. corresponding marked as unidentified. Therefore, it may have warning sign has been sized within an ignore region marked in the appeared next to a TestKitSuite and has not been reported fragment in the image. 2200 TapeStation...
Page 64
Troubleshooting Troubleshooting the Test Kit Manufacturers TestKitSuite.xml File 2200 TapeStation...
The purchase of this product conveys to the buyer the non-transferable right to use the purchased amount of the product and components of the product in research conducted by the buyer in conjunction with Agilent Technologies UK Limited’s automated gel electrophoresis system, where such research does not include testing, analysis or screening services for any third party in return for compensation on a per test basis.
Sound Emission Directive of 18 January 1991. This product has a sound pressure emission (at the operator position) < 70 dB. • Sound Pressure Lp < 70 dB (A) • At Operator Position • Normal Operation • According to ISO 7779:1988/EN 27779/1991 (Type Test) 2200 TapeStation...
With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a Monitoring and Control Instrumentation product. Do not dispose off in domestic household waste N O T E To return unwanted products, contact your local Agilent office, or see www.agilent.com more information. 2200 TapeStation...
Appendix Technical Service Technical Service For more information, please contact Agilent Technologies UK Limited www.agilent.com/genomics/contact 2200 TapeStation...
Page 70
TapeStation WEEE directive electronic waste shutdown procedure software error messages intended use start up sound emission specifications limitations load screentape TapeStation tips abort complete set up needle start change 2200 TapeStation...