Agilent Technologies 2200 TapeStation User Manual page 58

Hide thumbs Also See for 2200 TapeStation:
Table of Contents

Advertisement

6
Troubleshooting
Troubleshooting 2200 TapeStation ScreenPlex Images
Table 8
Overview troubleshooting 2200 TapeStation images
Image
Issue
The bands look
smudged – especially
towards the top of
the lane.
All of the sample
and/or ladder bands
appear very faint.
58
Probable cause
Solution
The sample has not
Always vortex samples for 5 seconds then spin
been mixed properly
down before placing in the 2200 TapeStation
with the loading
buffer.
The lane has not
Always spin down samples before placing in the
loaded optimally.
2200 TapeStation to remove any bubbles.
Very rarely, the 2200 TapeStation does not load
effectively - run the sample again.
2200 TapeStation