Frequency
Tools required
Parts required
CA UT IO N
Before beginning this procedure:
❏ Remove the top cover (page 128).
❏ Remove the transport assembly (page 132).
❏ Remove the sampling unit (page 134).
2 Remove the two fixing screws, and lift out the
board.
Repairing the Autosampler

SUD Board

SUD Board
When defective.
Pozidrive No. 1 screwdriver.
SUD board G1313-66503.
Take care not to damage the flex board when removing the SUD board.
Fixing screws
152
1 Disconnect all connectors from the SUD
board.
3 Install the new board. Ensure the board is
positioned between the board guide.