Frequency
Tools required
Parts required
Before beginning this procedure:
❏ Remove the top cover (page 128).
❏ Remove the transport assembly (page 132).
❏ Remove the sampling unit (page 134).
2 Feed the cable of the new sensor through the
base of the sampling unit to the SUD board.
Repairing the Autosampler

Leak Sensor

Leak Sensor
When defective
None.
Leak sensor 5061-3356.
1 Disconnect the leak sensor from the SUD
board (labelled "leak sensor"). Remove the
sensor.
3 Connect the connector to the connector
(labelled "leak sensor").
157