Frequency
Tools required
Parts required
Before beginning this procedure:
❏ Remove the front cover.
❏ Remove the vial tray.
❏ Remove the top cover (page 128).
❏ Remove the transport assembly (page 132).
Repairing the Autosampler

Sampling Unit

Sampling Unit
When accessing internal parts, or when defective.
1/4 inch wrench (supplied in accessory kit). Flat-head screwdriver.
Sampling unit G1313-60008 for G1313A
Sampling unit G1329-60008 for G1329A
Sampling unit G1329-60018 for G1389A
Sampling unit G2260-60008 for G2260A
The sampling units come without injection valve and analytical head assembly (see
"Sampling Unit Assembly" on page 162).
134
1 Remove the pump and column thermostat
capillaries (ports 1 and 6) from the injection
valve.