Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Safety Flap Missing

Safety Flap Missing
The safety flap was not detected.
Before the needle moves down into the needle seat to inject sample, the
safety flap locks into position. Next, and the gripper checks the safety flap by
trying to move the safety flap away from the needle. If the gripper is able to
move beyond the safety flap position (safety flap not in position), the error
message is generated.
• Safety flap missing or broken.
❏ Exchange the safety flap.