Frequency
Tools required
Parts required
CA UT IO N
Before beginning this procedure:
❏ Remove the top cover (page 128).
❏ Remove the transport assembly (page 132).
❏ Remove the sampling unit (page 134).
❏ Remove the ASM board (page 148).
❏ Remove the fan (page 146).
2Remove the power switch light pipe from the
coupler.
Coupler
Light
pipe
CellFrame
Repairing the Autosampler

Power Supply

Power Supply
When defective
Pozidrive No. 1 screwdriver
Power supply 0950-2528.
The ASM board is sensitive to electrostatic discharge. Always use the ESD
strap (see "Using the ESD Strap" on page 105) when handling electronic
boards.
154
1 Remove the bottom foam.
CellFrame
3Remove the two screws from the rear panel.