Table 47
Theory of Operation

Optional Interface Boards

Optional Interface Boards
The Agilent 1100 Series modules have one optional board slot that allows
addition of an interface board to the modules.
Optional Interface Boards
'HVFULSWLRQ
BCD Board
Fuse 250 mA (four are on the board)
LAN Board (see next page for details)
BCD Board
The BCD board provides a BCD output for the bottle number of the
Agilent 1100 Series autosampler and four external contacts. The external
contact closure contacts are relay contacts. The maximum settings are:
30 V (AC/DC); 250 mA (fused). There are general purpose cables available
to connect the BCD output, see "BCD Cables" on page 195 and the external
outputs, see "External Contact Cable" on page 198 to external devices.
228
3DUW 1XPEHU
G1351-68701
2110-0004