Frequency
Tools required
Parts required
CA UT IO N
N O T E
Before beginning this procedure:
❏ Remove the top cover and foam (page 128).
❏ Remove the transport assembly (page 132).
Repairing the Autosampler

ASM Board

ASM Board
When defective
5 mm wrench (for remote-connector screws).
7 mm wrench (for GPIB connector screws).
ASM board G1313-69520 for G1313A
ASM board G1329-69500 for G1329A - G1389A - G2260A
The ASM board is sensitive to electrostatic discharge. Always use the ESD
strap (see "Using the ESD Strap" on page 105) when handling electronic
boards.
This procedure requires reloading the autosampler firmware, reprogramming
of the instrument serial number, and realignment of the gripper.
148
1 Remove the connector screws from the GPIB
and RS232 connectors.
CellFrame
RS232
screws
GPIB
screws