Frequency
Tools required
Parts required
CA UT IO N
Before beginning this procedure:
❏ Remove the front cover.
Repairing the Autosampler
Rotor Seal
Rotor Seal
Poor injection-volume reproducibility
Leaking injection valve
1/4 inch wrench (supplied in accessory kit).
Hex key, 9/64 inch (supplied in accessory kit).
Rotor seal 0100-1853 (Vespel) for G1313A and G1329A
Rotor seal 0100-1849 (Tefzel) for G1313A and G1329A
Rotor seal 0100-2088 (Vespel) for G1389A
Rotor seal 0101-1268 (PEEK) for G2260A
The stator face is held in place by the stator head. When removing the stator
head, ensure the stator face does not fall out of the valve.
1 Remove all capillary fittings from the
injection-valve ports.
117