Sychronization Lost - Agilent Technologies G1313A Reference Manual

1100 series standard, micro and preparative autosamplers
Table of Contents

Advertisement

Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Sychronization Lost

Sychronization Lost
During an analysis, the internal synchronization or communication between
one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one
or more of the modules is no longer recognized as being connected to the
system, the error message is generated.
• CAN cable disconnected.
• Defective CAN cable.
• Defective main board in another module.
❏ Ensure all the CAN cables are connected correctly.
❏ Switch off the system. Restart the system, and determine which module or
modules are not recognized by the system.
❏ Ensure all CAN cables are installed correctly.
57

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1329aG1389aG2260a

Table of Contents