Injection-Valve Assembly - Agilent Technologies G1313A Reference Manual

1100 series standard, micro and preparative autosamplers
Table of Contents

Advertisement

Frequency
Tools required
Parts required
Before beginning this procedure:
❏ Remove all capillaries from the injection
valve (page 31).
❏ Remove the top cover (page 128).
❏ Remove the transport assembly (page 132).
❏ Remove the sampling unit (page 134).
2 Remove the three fixing screws.
Repairing the Autosampler

Injection-Valve Assembly

Injection-Valve Assembly
When defective
Pozidrive No. 1 screwdriver
1/4 inch wrench
Injection valve 0101-0921 for G1313A G1329A
Injection valve 0101-1050 for G1389A
Injection valve 0101-1267 for G2260A
140
1 Disconnect the valve-cable connector from
the SUD board (labelled "Valve Assy").
3 Remove the injection-valve assembly.