Compensation Sensor Open

Compensation Sensor Open
The ambient-compensation sensor (NTC) on the ASM board in the
autosampler has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
ASM board is dependent on ambient temperature. The change in resistance is
used by the leak circuit to compensate for ambient temperature changes. If
the resistance across the sensor increases above the upper limit, the error
message is generated.
• Defective ASM board.
❏ Exchange the ASM board.