Exchanging the Power Supply Fuses
The fuse holders are located on the rear panel of the autosampler thermostat.
When
If wrong fuses are installed. Required is 2.5A fuses.
Tools required
• Flat head screwdriver
Parts required
#
2110-0015
Damaged electronics
WA R N I N G
Disconnecting or reconnecting the autosampler to autosampler thermostat cable
when the power cords are connected to either of the two modules will damage the
electronics of the modules. In such a case, mainboards of both instruments must be
exchanged, otherwise they can damage the other instrument.
➔
1 Switch OFF the power switch at the front of the thermostatted autosampler.
2 Remove the power cable from the two modules.
1200 Series AT User Manual
Description
Fuses T2.5 A/250V (CSA, UL listed)
Make sure the power cords are unplugged before disconnecting or reconnecting the
autosampler to autosampler thermostat cable.
Introduction into Repairing the Thermostat
5
Maintenance
49