Injection-Valve Assembly - Agilent Technologies 1200 Series Reference Manual

Standard and preparative autosamplers
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Injection-Valve Assembly

Frequency
Tools required
Parts required
Preparations for this
procedure
1 Disconnect the valve-cable connector from the SUD board
(labelled "Valve Assy").
Agilent 1200 Series Autosampler Reference Manual
When defective
Pozidrive No. 1 screwdriver
1/4 inch wrench
Injection valve 0101-0921 for G1329A
Injection valve 0101-1267 for G2260A
Remove all capillaries from the injection valve
Remove the top cover
(page
Remove the transport assembly
Remove the sampling unit
2 Remove the three fixing screws.
Repairing the Autosampler
(page
134).
(page
139).
(page
141).
4
35).
147