Injection-Valve Assembly; Removing The Injection Valve Assembly - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Injection-Valve Assembly

When required
Tools required
Parts required

Removing the injection valve assembly

1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the sampling unit, (see
7 Disconnect the valve-cable connector from the SUD board (Valve Assy).
8 Remove the three screws which fix the injection valve assembly.
9 Remove the injection valve assembly.
1100 Series WS MWS Reference Manual
When defective
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
0101-0921 Injection Valve assembly for the G1367A/68A samplers
0101-1050 Micro Injection valve assembly for the G1377A/78A samplers
Needle/Seat" function.
page 132).
page 133).
Repairing the Sampler
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the sampling unit"
4
on
on
on page 134).
139

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1377a

Table of Contents