Trackmaster FVX325 Field Service Manual page 42

Table of Contents

Advertisement

TRACKMASTER
5.
P 5 will appear in the Center Display with default setting of 15.0 in the Incline
Window. Using the numeric keypad, change to 25.00 for medical units. For fitness
units use default of 15.0. Press SELECT key.
Center Display Action required
P 5
Enter Maximum Grade
needed INCLINE arrow
keys
6.
P 6 will appear in the Center Display with default setting of 4 in the Speed Window.
Do Not Change This Number. Press SELECT key.
7.
P 7 will appear in the Incline Display with default setting of 3175 n the Center
Display. Do not change this Number unless instructed to by the manufacturer. Press
SELECT key.
8.
P 8 will appear in the Center Display. This setting used when the treadmill has
the Contact Heart-rate feature. If the treadmill only has the Polar ® Heart-rate
feature, use INCLINE arrows to change setting to POLAR H-RATE. Press
SELECT key.
Center Display
Action Required
P 8
Select Serial Heart-rate
or Polar Heart-rate by
using INCLINE arrow
keys
9.
P 9 will appear in the Center Display. Use the up Incline key to change to Normal
SCO. Medgraph SCO setting works only when the treadmill is used in conjunction
with a Medgraphics® Stress system. Press SELECT key.
Center Display
Action Required
P 9
Select Medgraph
SCO or Normal
SCO by using
incline arrow keys
10.
P 10 will appear in the Center Display. Use Incline down key to change to ON if
RS232 port is to be used. Otherwise, RS232 port should remain off. Press SELECT
key.
Controller
Action Required
Display
P 10
Select 232 Comm Off or
232 Comm On
by using incline arrow keys
August, 2006
Ct
Parameter Options
M Grade
15 for 15% elevation
fitness units
25 for 25% elevation
medical units
Changes Affect
Select combined
Polar & contact
heart rate or Polar
only
Changes Affect
Selection of Medgraph SCO
allows serial communication
with the Medgraphics®
systems only. Selection of
Normal SCO allows serial
communication with other
stress systems
Changes Affect
Selecting 232 Comm
Off cuts off RS232 port
Selecting 232 Comm On
restores RS232 port
communication
FVX, TMX Series Service Manual
Observe
15.0 appears in the
Center Display
Parameter
Observe
Options
Select Polar ®
Hr appears in the
H-rate for chest-
Incline Display, Pu
strap only
appears in the
Speed Display
Parameter
Observe
Options
Medgraphics
"SCO" appears
SCO
in the Incline
Display
Normal SCO
Parameter
Options
232 Comm Off
232 Comm On
Observe
"232" appears
in the Incline
Display
5-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tmx435Fvx325cTmx425cpFvx325cpTmx425c

Table of Contents