Making Measurements using Frequency Dependent Offset Tables
To view the frequency dependent offset tables currently stored in the
System
power meter, press
Tbls" screen is displayed as shown in Figure 2-6.
There are ten frequency dependent offset tables named CUSTOM_A
through CUSTOM_J which do not contain any data when the power meter
is shiped from the factory.
You cannot delete any of the 10 existing frequency dependent offset tables
or create any additional tables. However, you can edit the 10 existing
tables. Each frequency dependent offset table can contain a maximum of
80 frequency points.
To edit a table:
1. Select one using the
displayed tables. When the table you want to edit is highlighted
press
Edit Table
shown in Figure 2-7 on page 2-33. In this screen you can modify,
add and remove frequencies and offset and also edit the table
name. Use the
between the table name and its frequencies and offsets.
A frequency in the range of 0.001 MHz to 999.999 GHz can be
entered.
An offset in the range of 1% to 150% can be entered.
The following rules apply to naming frequency dependent offset
tables:
The name must consist of no more than 12 characters.
Agilent E4419B User's Guide
,
,
Tables Freq Dep Offset
Inputs
Figure 2-7: "Edit" Screen
and
. The "Edit Offset" screen is displayed as
,
,
and
Power Meter Operation
. The "Offset
hardkeys to scroll through the
hardkeys to move
2-33