Step Detection - Agilent Technologies E4419B User Manual

Dual channel power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Step Detection

To reduce the filter settling time after a significant step in the measured
power the filter can be set to re-initialise upon detection of a step
increase/decrease in the measured power. Step detection can be set in both
manual and automatic filter modes.
To set step detection for channel A:
System
1. Press
Inputs
2. Press the
ChA Filter
3. Use the
Step Det Off On
detection.
For channel B use the steps above except use
.
ChB Filter
Agilent E4419B User's Guide
,
A Input Settings
softkey to access the filter menu.
softkey to enable /disable step
Power Meter Operation
Setting Averaging
,
.
More
B Input Settings
and
2-43