Menu Reference
Front Panel Menu Reference "Zero/Cal" Menu
Zero
Cal
The softkey menu structured to this hardkey affects the channel setup.
Press this hardkey to access the "Zero/Cal" menu which allows you to zero
and calibrate the power meter. Refer to "Calibrating the Power Meter", on
page 2-8 for further information.
Zero A
Press this softkey to adjust channel A for a zero power reading
with no power applied to the power sensor. Refer to "Zeroing and
Calibrating the Power Meter", on page 2-7 for further information.
Zero B
Press this softkey to adjust channel B's internal circuitry for a
zero power reading with no power applied to the power sensor.
Refer to "Zeroing and Calibrating the Power Meter", on page 2-7
for further information.
Zero Both
Press this softkey to adjust both channel A's and channel B's
internal circuitry for a zero power reading with no power supplied
to the power sensor. Refer to "Zeroing and Calibrating the Power
Meter", on page 2-7 for further information.
Cal
Press this softkey to access a menu which allows you to calibrate
the power meter. Calibration sets the gain of the power meter
using a 50 MHz oscillator as a traceable power reference.
Cal A
Press this softkey to calibrate channel A with the connected
power sensor. The POWER REF output is used as the signal
source for calibration and is automatically switched on during
this procedure.
A Ref CF
Press this softkey to enter a reference calibration factor for
channel A. A value in the range of 1% to 150% can be entered.
The default value is obtained from the sensor calibration table
if one is selected, otherwise it is 100%. Use the
and
choice press
using an Agilent E-series power sensor.
3-34
hardkeys to change the value. To confirm your
%
. This softkey cannot be selected when you are
,
Agilent E4419B User's Guide
,