Power Meter Operation
Making Measurements using Frequency Dependent Offset Tables
Note
Note
2-34
All characters must be upper or lower case alphabetic
characters, or numeric (0-9), or an underscore (_).
No other characters are allowed.
No spaces are allowed in the name.
2. To edit the parameter you currently have selected press
The power meter displays the data in a pop up window. Modify
this data (see below) as desired.
Use
or
is currently positioned.
Use
or
3. To confirm your choice press the appropriate softkey.
4. To adda new frequency dependent offset table entry, press
. You are prompted for the frequency and offset. The entry
Insert
is sorted by frequency.
5. To remove a frequency dependent offset table entry, use the
,
and
you delete the frequency the equivalent offset is also removed and
vice versa.
You can only edit the table name you cannot delete it.
6. When you have completed editing the frequency dependent offset
table, press
Done
Ensure that the frequency points you use cover the frequency
range of the signals you want to measure. If you measure a signal
with a frequency outside the frequency range defined in the
frequency dependent offset table, the power meter uses the
highest or lowest frequency point in the frequency dependent
offset table to calculate the offset.
to modify the character on which the cursor
to move to other characters.
hardkeys to select the entry, press
to return to the "Offset Tbls" screen.
Change
Delete
Agilent E4419B User's Guide
.
,