Figure 2-14: Remote I/O TTL Outputs
1 2 3 4 5 6 7 8
To set the limits:
Meas
1. Press
Setup
minimum limits are displayed under the
respectively.
2. To change either of these settings press the appropriate softkey.
The power meter displays the current value in a pop up window.
Modify this value (see below) until the desired value is displayed.
Use
currently positioned.
Use
3. To confirm your choice press the appropriate measurement units.
Limits can be disabled and re-enabled simply by pressing
Limits Off On
Agilent E4419B User's Guide
Pin Number
1
2
3
4
5
6
7
8
,
. The current setting of the maximum and
Limits
or
to modify the digit on which the cursor is
or
to move to other digits.
.
Power Meter Operation
Setting Measurement Limits
Connection
none
Ground
Upper Window TTL Output
Lower Window TTL Output
TTL Input 1
TTL Input 2
Ground
Ground
and
softkeys
Max
Min
2-49