Menu Reference
Front Panel Menu Reference "Frequency/Cal Fac" Menu
A Cal Fac
Press this softkey to enter the measurement calibration factor for
channel A in the range of 1% to 150%. The default value is 100%.
Use the
calibration factor. To confirm your choice press
calibration factor using this softkey is only possible when you are
using an Agilent 8480 series power sensor.
B Freq
Press this softkey to enter the frequency of the signal to be
measured on channel B, in the range of 0.001 MHz to
999.999 GHz. The default value is 50 MHz. Use the
and
choice press the appropriate frequency unit. Entering a frequency
using this softkey is only possible when you have a sensor
calibration table selected or you are using an Agilent E-series
power sensor.
B Cal Fac
Press this softkey to enter the measurement calibration factor for
channel B in the range of 1% to 150%. The default value is 100%.
Use the
calibration factor. To confirm your choice press
calibration factor using this softkey is only possible when you are
using an Agilent 8480 series power sensor.
3-14
,
,
and
hardkeys to change the frequency. To confirm your
,
,
and
hardkeys to change the
. Entering a
%
hardkeys to change the
. Entering a
%
Agilent E4419B User's Guide
,
,