Menu Reference
Front Panel Menu Reference "More", "Preset/Local", "Prev" Keys
More
Press this hardkey to move through all the softkeys available on a
particular level of a menu. The bottom right of the power meter display
indicates the number of pages in the menu. For example, if "1 of 2" is
displayed, pressing
More
moves you back to "1 of 2".
Preset
Local
Press this hardkey to preset the power meter if you are currently working
in local mode (that is, front panel operation). A confirmation pop up
window is displayed prior to a preset being carried out. Presetting the
power meter returns you to the "Contrast" menu. However, if you are in
remote mode (that is, GP-IB operation), then pressing this hardkey places
the power meter in local mode provided local lock out (LLO) is not enabled.
When reverting to local mode the power meter triggering is set to free run.
Prev
Press this hardkey to move back one level in the softkey menu structure.
Pressing this hardkey numerous times eventually returns you to the
"Contrast" menu which allows you to increase or decrease the display
contrast.
•
Press this softkey to increase the display contrast.
•
Press this softkey to decrease the display contrast.
3-18
moves you to "2 of 2". Pressing
again
More
Agilent E4419B User's Guide