Making Measurements with the Agilent E-Series Power Sensors
10. You must set the display to the type of measurement you require.
This can either be a direct channel measurement, a ratio
measurement of both channels, or a difference measurement
between both channels. Use the following key presses:
Meas
Press
,
Setup
the measurement you require, either
then
, or
A-B
11. The measurement result is now displayed.
Example
To make a measurement on channel A using an Agilent E-series power
sensor. The frequency of the signal to be measured is 100 MHz.
Disconnect the power sensor from any source.
Zero
Press
.
Cal
Press
Zero A
Connect the power sensor to the POWER REF output.
Press
,
Cal Cal A
Frequency
Press
Cal Fac
hardkeys to enter 100. Press
Connect the power sensor to the signal to be measured.
The measurement result is now displayed.
Agilent E4419B User's Guide
. Select the appropriate softkey for
Input Select
.
B-A
.
.
,
. Use the
A Freq
.
MHz
Power Meter Operation
A B A/B B/A
,
,
,
, or
,
,
and
,
More
2-17