Initial Operation - Ahlborn ALMEMO 6290-7B Operating Instructions Manual

Measuring instrument with built-in printer
Table of Contents

Advertisement

2. INITIAL OPERATION

1. Connect the transducers to the ALMEMO
2. For battery charging connect mains adapter ZB6290-NA to socket (9), s.3.2.
3. Use key ON/OFF to switch the instrument on, see 3.4.
Attention! The keys might be locked by the key switch (8)!
4. Displaying Measured Values
Select function MEASURE by using the key MEASURE,
select measuring channel by using the key CH▲, read meas. value, s.7.1.1.
Scan all measuring points automatically by using key START/STOP, see 7.2.1.
5. Print Measured Values
Select function output mode ´OM´ by using the key FUNCTION.
Set output mode ´dr´ by using the keys ENTER, "#,$, see 8.1.
Select function print cycle ´PC´ by using the key FUNCTION.
Enter print cycle, see 7.2.3.
Enter correct time and date, as required, see 7.2.4.
Start/stop measurement by using the key START/STOP, see 8.3.2.
6. Plot Measured Values
Set output mode ´PL 1´ or ´PL 2´ in function ´OM´, see 8.1, 8.4.
Select the functions ´PS´ and ´PE´ successively by using the key FUNCTION
and enter the plot range start and range end for all plot channels (s. 8.4.1).
Select function paper feed ´PF´ by using the key FUNCTION,
enter paper feed in mm/h, see 8.4.2.
Start/stop measurement by using the key START/STOP, see 8.4.3.
7. Memory Output to Printer or Computer (with option S only)
Connect peripheral device, via interface cable, to socket A1, see man. 5.2.
Set peripheral device to 9600 bd, 8 data bits, 1 stop bit, no parity.
Enter output mode ´U´ in function ´OM´ (see 8.1) and use the key CH▲
for setting the output format ´n´ or spreadsheet (table) ´t´, see 8.5.2.
Select function MEMORY ´MY´ by using the key MEASURE
Start memory output by using the key START/STOP, see 8.6.2.
Clear the memory in function ´MY´ by using the keys ENTER, Clr, see 8.6.1.
8. Display Max and Min Values
by using the key MEASURE in function ´MH´ or ´ML´, see 7.1.2.
sockets M1
ALMEMO
6290-7B
Initial Operation
and M2
/M3
/M4
(6), s. 4.
12

Advertisement

Table of Contents
loading

Table of Contents