Function RANGE ´R´
Selection with key:
Change meas. range:
Transducer
Pt100-1
Pt100-2
Ni100
NiCr-Ni (K)
NiCroSil-NiSil (N)
Fe-CuNi (L)
Fe-CuNi (J)
Cu-CuNi (U)
Cu-CuNi (T)
PtRh10-Pt (S)
PtRh13-Pt (R)
PtRh30-PtRh6 (B)
Au-FeCr
Ntc type N
Millivolt 1
Millivolt
Millivolt 2
Volt
Differential-Millivolt 1
Differential-Millivolt
Differential-Millivolt 2
Differential-Volt
Sensor Voltage
Milliampere
Percent (4-20mA)
Ohm
Frequency
Pulses
Digital input
Digital interface
Infrared 1
Infrared 2
FUNCTION
...
Example: channel M1, range NiCr, dimension °C
ENTER
,
... or
Connector /
Cable / Sensor
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9020-FS
ZA 9020-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9000-FS
ZA 9050-FS
ZA 9050-FS
ZA 9050-FS
ZA 9050-FS
ZA 9000-FS
ZA 9601-FS
ZA 9000-FS
ZA 9000-FS
ZA 9909-AK
ZA 9909-AK
ZA 9000-EK2
ZA 9919-AKxx
ZA 9000-FS
ZA 9000-FS
ALMEMO
Sensor Programming
1: N i C r
ENTER
... ,
Meas. Range
-200.0... +850.0
-200.00...+200.00
-60.0... +240.0
-200.0...+1370.0
-200.0...+1300.0
-200.0... +900.0
-200.0...+1000.0
-200.0... +600.0
-200.0... +400.0
0.0...+1760.0
0.0...+1760.0
+400.0...+1800.0
-270.0...
+60.0
-30.00...+125.00
-26.000...+26.000
-10.000...+55.000
-260.00...+260.00
-2.6000...+2.6000
-26.000...+26.000
-10.000...+55.000
-260.00...+260.00
-2.6000...+2.6000
0.00...20.00
-32.000...+32.000
0.00... 100.00
0.00... 400.00
0... 25000
0... 65000
0.0... 100.0
-65000... +65000
0.0... +200.0
0.0... +800.0
6290-7B
°C
.
Display
Dim
P104
°C
P204
°C
N104
°C
NiCr
°C
NiSi
°C
FECO
°C
IrCo
°C
CUCO
°C
CoCo
°C
Pt10
°C
Pt13
°C
EL18
°C
AUFE
°C
Ntc
°C
U 26
mV
U 55
mV
U260
mV
U2.60
V
d 26
mV
d 55
mV
d260
mV
d2.60
V
UbAt
V
I032
mA
P420
%
Ohn
FrEq
Hz
PULS
Inp
%
diGi
Ir 1
°C
Ir 2
°C
22

Advertisement

Table of Contents
loading

Table of Contents