Limit Values; Correction Values, Sensor Adjustment - Ahlborn ALMEMO 6290-7B Operating Instructions Manual

Measuring instrument with built-in printer
Table of Contents

Advertisement

Sensor Programming

6.4 Limit Values

Two limit values (MAX and MIN) can be programmed for each measuring
channel. The exceeding of the limit values is handled as a fault, similar to the
exceeding of the measuring range limits and sensor breakage. The arrow
ALARM will appear in the display and the alarm relays will respond (see
manual 6.3.9). An exceeding can also be used to start or stop a measuring
point scan (see manual 6.6.3).
Function LIMIT VALUE HIGH ´LH´ and LIMIT VALUE LOW ´LL´
Selection with key:
Programming:
Switch-off:
6.5 Correction Values
The correction values ZERO POINT and SLOPE allow for correcting sensors
with regard to zero point and slope (gain) (see manual 6.3.10).
Corrected Meas. Value = (Meas. Value - ZERO POINT) x SLOPE.
Function ZERO POINT CORRECTION ´ZC´
Selection with key:
Programming:
Clear:
Function SLOPE CORRECTION (GAIN): ´SC´
Selection with key:
Programming:
If correction values are programmed and, as a result, the measured value is
corrected, the arrow CORR will be indicated in the display.
25
FUNCTION
Input according to 5.4
ENTER
FUNCTION
Input according to 5.4
ENTER
FUNCTION
Input according to 5.4
ALMEMO
6290-7B
...
Clr
,
...
Clr
,
...
1: 1 2 3.0
1: - - - -
1: 0 0 3.2
1: - - - -
1:1.5 0 0 0
LH
LH
ZC
ZC
SC