Atmospheric Pressure Compensation - Ahlborn ALMEMO 6290-7B Operating Instructions Manual

Measuring instrument with built-in printer
Table of Contents

Advertisement

Zero Point Adjustment
Many sensors must be adjusted at least once or at regular intervals to
compensate for instabilities. For this purpose, a specific zero point
adjustment is available, in addition to the ´Set Measured Value to Zero´
mentioned above, as some sensors require an additional scaling (e.g. pH
probes). In this function the zero point error is not stored as base value but as
zero point correction (special cases and slope correction, see 6.5). In this case,
the locking mode must be set below 4 (see 6.7). The zero point correction is
performed using the following keys:
Function MEASURE:
If a base value is programmed the measured value is not indicated
as zero but as the negative base value after the adjustment.

7.1.4 Atmospheric Pressure Compensation

Some measuring variables depend on the environmental atmospheric pressure
(see 6.2 measuring range list ´w. PC´). As a result, higher deviations from the
normal pressure of 1013mbar can cause corresponding measuring errors:
e.g. error per 100mbar:
Therefore, the atmospheric pressure should be considered (approx.
-11mb/100m over mean sea level, MSL) especially during use in a
corresponding height above sea level. It can either be programmed or
measured with a sensor (see manual 6.7.2).
Function ATMOSPHERIC PRESSURE ´mb´
Selection with key:
With each reset the atmospheric pressure is set to 1013mb. It can be set to the
actual value by the usual data entry.
Continuous Measurement of a Measuring Point
MEASURE
rel. humidity psychrometer
dynamic pressure
O
saturation
2
FUNCTION
... Input mbar, see 5.4
ALMEMO
Zero point adjustment:
6290-7B
ENTER
,
approx. 2%
approx. 5%
approx. 10%
1 0 1 3
m b
30

Advertisement

Table of Contents
loading

Table of Contents