OPERATION

4.6 Imbalance Detection

With fast filling of the basket in the acceleration phase, an imbalance can occur. In this case, the
centrifuge switches off the drive immediately and the basked decelerates to standstill.
The display "time" (A-3) will show the word "error" (M11) and the number "01" appears.
When inside the display "time" (A-3), the word "error" along with the number "02" appear, (see
Figure 23) the cause may be as follows. -> The imbalance switch is defective.
Figure 23

4.7 Volume Flow

The volume flow of the medium to be analysed may be max. 3.5 l/min. When the flow rate is higher,
the temperature of the drive increases until it is switched off automatically for safety reasons.

4.8 External Temperature Control

The Sieva-3, with the preparation for external temperature equalization, is equipped with a copper
pipe winding around the centrifuge chamber. On the side of the housing there are screw connections
with an internal thread of ¼''. Hence, nipples of various sized can be used.
The tempering liquid may have a max. temperature of 90°C. The liquid may be pumped through the
cooper pipe winding with a max. pressure of 5 bar.