Input Type: (Dc Voltage/Process Value) - Omega DP41-B User Manual

Universal input meter
Hide thumbs Also See for DP41-B:
Table of Contents

Advertisement

12
Input Type Selection

12.3 INPUT TYPE: (DC VOLTAGE/PROCESS VALUE)

Press 'MENU'
Press 'RESET/ENTER'
Press ' /MAX'
VOLt
Press ' /MAX'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press ' /MAX'
Press 'RESET/ENTER'
Press ' /MAX'
Press 'RESET/ENTER'
Display
UN IPOL
100 MV
1 V
10 V
100 V
bI POL
50 MV
500 MV
5 V
50 V
46
ENTER INPUT TYPE MENU:
1) Display shows "INPUt" Input Menu.
2) Display flashes "VOLt" (default) or previous stored input type.
3) Scroll through the available following input type selection list. Display
flashes:
"VOLt" (DC Voltage)
"CURRNt" (DC Current)
"bRIdGE" (Bridge)
"POt" (Potentiometer)
"tC" (Thermocouple)
"Rtd" (RTD)
If the displayed input type is "VOLt", press 'RESET/ENTER' to skip to Step 6
("VOLt" stops flashing).
"VOLt" DC VOLTAGE SUBMENU:
4) Scroll through the available selection to "VOLt" (flashing).
5) Display shows "VOLt" again, but not flashing.
6) Display flashes "UN IPOL" unipolar (default) or previously stored
selection of voltage input type "bI POL" bipolar.
You must choose bi-Polar or uni-Polar before selecting Bridge or
Potentiometer
7) Scroll through the 2 available voltage input types to the selection of
your choice.
8) Display flashes "100 MV"(default) or previous stored input voltage
range selection. If the display matches your choice of Voltage range. Go
to Step 10.
9) Scroll through the available voltage ranges to the selection of your
choice (see below Table for Voltage Input types/Ranges selection).
10) Display shows "StOREd" stored message momentarily and then
advances to the "RdG.CNF" Reading Configuration Menu.
Voltage types
unipolar
unipolar
unipolar
unipolar
bipolar
bipolar
bipolar
bipolar
Voltage Ranges
100 milliVolt DC
1 Volt DC
10 Volt DC
100 Volt DC
± 50 milliVolt DC
± 500 milliVolt DC
± 5 Volt DC
± 50 Volt DC