Multi-Point Scale and Offset Menu Configuration
ON.L INE
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press ' /MAX' & '
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press 'RESET/ENTER'
Press ' /MAX' & '
Press 'RESET/ENTER'
MANUAL
To scale "MANUAL", calculate inputs based on transducer specifications and manually enter them via
on the front panel push buttons. The following example assumes load cells with this specification:
Maximum Load:....................................................................100.0 lb
Output ................................................................................3.0 mV/V
Sensor Excitation ......................................................................10 V
Maximum Sensor Output = 3.0(mV/V) x 10(V) = 30mV
IF "ON.L INE" and "NUM.PNt" = 01 are YOUR CHOICE:
(On-Line Calibration with 2 points)
6) Display stops flashing "ON.L INE" On-line Calibration or Scaling.
7) Display shows "INPUt 0".
8) Display shows the actual live-input signal being received and equal to
approximately the minimum input signal or 0% of known load range.
9) Display shows "StOREd" momentarily and advances to "REAd 0".
10) Display shows last stored value of "REAd 0" with 1st digit flashing.
/MIN'
11) To enter the value of "REAd 0".
This value corresponds to "INPUt 0" in terms of some meaningful
engineering units. To show "INPUt 0" as minimum or 0% as enter
"REAd 0" value = 000000 (Example)
12) Display shows "INPUt 1".
13) Display shows the 2nd higher live-input signal being connected for
your preference as equal to approximately 100% of the transducer range
or the max. Input signal.
14) Display shows "StOREd" momentarily and advances to "REAd 1".
15) Display shows last stored value of "REAd 1" with 1st digit flashing.
/MIN'
16) To enter the value of "REAd 1". This value corresponds to "INPUt 1"
in term of some meaningful engineering units. To show "INPUt 1" as
maximum or 100% as enter "REAd 1" value = 000100 (example).
17) Display shows "StOREd" momentarily and advances to "OUt.CNt"
Output configuration menu.
IF "MANUAL" and "NUM.PNt"=01 are YOUR CHOICE
(Scaling without Known Loads or based on specifications)