Operating Requirements

Electrical Ratings

Power Requirements
WVR7200 Installation and Safety Instructions
This section provides the specifications that you need to know to operate your
product safely and correctly. Refer to the complete product specifications for
additional information.
Power connector
The instrument has the following power requirements:
A single-phase power source with one current-carrying conductor at or near
earth-ground (the neutral conductor).
The power source frequency must be 50 or 60 Hz, and a operating voltage
range must be from 100 to 240 VAC, continuous.
WARNING.
To reduce risk of fire and shock, ensure the mains supply voltage
fluctuations do not exceed 10% of the operating voltage range.
Systems with both current-carrying conductors live with respect to ground
(such as phase-to-phase in multiphase systems) are not recommended as
power sources.
NOTE.
Only the line conductor is fused for over-current protection. The fuse is
internal and not user replaceable. Do not attempt to replace the fuse. If you
suspect the fuse has blown, return the unit to an authorized service center for
repair.
Operating Requirements
7