Agilent Technologies E4980A Service Manual page 33

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Figure 4-3
SELF TEST Page
Choosing a Test Item
Functional Description E4980A can run the following self tests:
The tests of SYSTEM, USER DATA, and BATTERY are also performed by the power on
test.
Table 4-2
Test item
SYSTEM
USER DATA
BATTERY
KEY
DISPLAY
BIAS INTERFACE
Chapter 4
Troubleshooting Using Service Function
Description
Checks the entire system, A1/A2/A3 boards, system
calibration data, and so on.
Checks the GPIB/LAN configurations, instrument
configuration information, instrument calibration data, and
scanner calibration data.
Checks the internal batteries.
Checks the font panel keys. (Visual checks only; no result
shown on screen.)
Checks the font panel LEDs and LCD. (Visual checks only; no
result shown on screen.)
Checks the bias interface. (Visual checks only; no result shown
on screen.)