Troubleshooting
Performance test failure troubleshooting

Adjustment failure troubleshooting

Table 4-14 represents the relationships between the failed adjustment and probably faulty
assembly. Note that this table lists some typical cases. There are possibilities that another
assembly may be faulty.
Table 4-14
Adjustment failure troubleshooting information
Adj
First failed adjustment
Num.
1
Null DC offset adjustment
2
Null ADC offset adjustment
3
ADC Linearity adjustment
4
DC Bias adjustment
5
DC Level V Monitor adjustment
6
DC Level I Monitor adjustment
7
DC Source adjustment
8
DCR adjustment
9
Impedance adjustment
10
AC level Monitor adjustment
11
OSC Level adjustment
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
58
Probable faulty board assembly