Performance Test Failure Troubleshooting - Agilent Technologies E4980A Service Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Performance test failure troubleshooting

Table 4-15 represents the relationships between the failed test and probable faulty
assembly. If the performance test failure cannot be removed by a proper adjustment,
replace the assembly shown in this table.
Note that this table lists some typical cases. There are possibilities that another assembly
may be faulty.
Table 4-15
Performance test failure troubleshooting information
Test
First failed test
Num.
1
Frequency Accuracy Test
2
DC Source Level Accuracy Test
3
DC Bias Level Accuracy Test
4
DC Bias V Level Monitor Accuracy Test
5
DC Bias I Level Monitor Accuracy Test
6
AC Signal Level Monitor Accuracy Test
7
AC Signal Level Accuracy Test
8
Impedance Measurement Accuracy Test
9
DCR Measurement Accuracy Test
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
Chapter 4
Performance test failure troubleshooting
Probable faulty board
assembly
A1
A2
#
#
#
###
###
###
#
###
###
Troubleshooting
A3
PPMC
###
###
###
#
###
#
#
59