Correction data measurement is aborted.
1201
CPU Bd EEPROM write error
An error is generated while writing data to EEPROM. When this error occurs, contact
Agilent Technology's Sales and Service Office or the company from which you bought the
device.
1200
CPU Bd FLASH ROM write error
An error is generated while writing data to FLASH. When this error occurs, contact
Agilent Technology's Sales and Service Office or the company from which you bought the
device.
29
Cycle power of 42841A and Preset
The connection was not established with the 42841A. Cycle the power of the 42841A, then
preset the E4980A.
D
90
Data buffer overflow
This error occurs when the amount of data exceeds the defined data buffer memory size. If
this error occurs during the measurement after pressing [Save/Recall] - SAVE DATA -
START LOG, press SAVE & STOP.
-230
Data corrupt or stale
The setting file or correction data to be read cannot be read out with the current firmware
version because the setting file or correction data is corrupt or old.
-222
Data out of range
A data element (not violating the standard) outside the range defined by this instrument has
been received. This error occurs, for example, when an integer-based command for which
the parameter can be rounded exceeds the range of -2147483648 to +2147483647 or when
a real-number-based command for which the parameter can be rounded exceeds the range
of -9.9e37 to +9.9e37. This error also occurs when a numeric value other than a specified
one is entered into a command in which the "BIN number" and "list table number" are
specified as parameters and thus the parameters are not rounded.
-104
Data type error
Improper data type used (for example, string data was expected, but numeric data was
received).
26
DC bias I sink overload
DC bias sink current is overloaded. When this error occurs, lower the value of the current
setting.
25
DC bias I source overload
DC bias source current is overloaded. When this error occurs, lower the value of the
current setting
19
DC bias not available
DC bias cannot be turned on in Rdc measurement.
18
DC bias opt not installed
Appendix D
Error Messages
Error number: 1201
143