Agilent Technologies E4980A Service Manual page 108

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Replacement Procedure
Base Plate Replacement
NOTE
When you fasten the ground cable and the Base Plate with the two TORX T20 screws, set
screwdriver's torque to 21.1 in-lb.
When you fasten the Base Plate with the seven TORX T20 screws, set screwdriver's torque
to 21.1 in-lb.
When you fasten the Shield Angle with the two TORX T10 screws, set screwdriver's
torque to 8.7 in-lb.
Figure 6-10
Base Plate Board Removal